Product Features
The Diazyme HDL-Cholesterol assay is a direct dual liquid stable reagent system which can be used reliably with fasting and non-fasting specimens. Convenient instrument specific packaging options including Roche Hitachi 917 series, Olympus AU (400/600/640/680), Beckman Synchron CX, LX and DXC. The assay utilizes Diazyme’s proprietary PVS\PEGME coupled technology which provides outstanding performance. The assay features a linear range of 1.1 to 184.8 mg/dL, intra and inter %CV values <2.6 % and less than 10% interference from triglycerides to 1000 mg/dL, ascorbic acid to10 mmol/L, bilirubin to 40 mg/dL, bilirubin conjugate to 30 mg/dL, hemoglobin to 1000 mg/dL. Stable for 12 months from date of manufacturer when stored @ 2° - 8° C. 

Assay Principle
The assay is based on a modified polyvinyl sulfonic acid (PVS) and polyethylene-glycol-methyl ether (PEGME) coupled classic precipitation method with the improvements in using optimized quantities of PVS/PEGME and selected detergents. LDL, VLDL and chylomicron (CM) react with PVS and PEGME and the reaction results in inaccessibility of LDL, VLDL and CM by cholesterol oxidase (CHOD) and cholesterol esterase (CHER). The enzymes selectivity react with HDL to produce H2O2 which is detected through a Trinder reaction. 


Intended use
The Diazyme HDL-Cholesterol reagent is intended for the quantitative in vitro determination of High Density Lipoprotein Cholesterol in human serum or plasma. The reagent can assist in the diagnosis and treatment of patients at risk of developing coronary heart disease. Low HDL cholesterol is related to the high risk of coronary heart disease. 

Product             

Catalog Number             

Packaging             

Method/Format             

Kit (1120 Tests)

DZ129A-K

R1: 4 x 70 mL
R2: 2 x 50 mL

Modified PVS/PEGME/
Liquid Stable

Calibrator

DZ130A-Cal

1 x 1 mL

 

Control

DZ130A-Con

2 x 3 mL