  • ETI-MAX3000 Specifications.
     

    SAMPLE AND DISPENSING UNIT

    * Liquid handling                    1 Syringe of 1 ml capacity

    * Disposable tips                    Carbon, 300 or 1000 ㎕, Automatically managed by software

    * Pippetting area                    4 plate handling

    * Precision (Sample & Reagent)    CV<5.0% with 10 ㎕,  CV<1.5% with 100 ㎕

    * Level sensor system              Electronic

    * Clot detection                     Yes

    * Predilution Module                Above to 3.8ml (Total predilution volume)

    * Predilution  Rack                 Up to 160

    * Multidispensing                   Yes(sample, control & reagent)

    * Sample dispensing time           < 18 min/96 well(100㎕/well)

    * Reagent dispensing time          < 4 min/96well(100㎕/well)

    * Carryover                         None


    SAMPLE IDENTIFICATION UNIT

    * Identification                      Barcode scanner for primary tubes. controls &reagents

                                            Barcode scanner for microplate(optional)

                                          Manual barcode gun(optional, connected in emulation keyboard)

    * Tube                              11 - 12.5 mm diameter, 55 - 100 mm height

                                          16 mm diameter, 100 mm height

    * Labels                            Interleaved 2 of 5, UPC A & E, IATA 2 of 5,

                                          Industrial 2 of 5, EAN 8 or 13, Code 128,

                                          EAN 128, Pharmacode, EAN Addendum 2 or 5, Code-a-bar

    * Capacity                          Up to 240 sample tubes.


    INCUBATION UNIT

    * Capacity                          4 independent chambers

    * Temperature range                5 ℃ above room temperature to 50 ℃

    * Accuracy                        ± 0.7 ℃ mean of plate

    * Uniformity                         ± 0.7 ℃ across plate

    * Shaking                           Longitudinal.



    WASHING UNIT

    * Capacity                          Up to 4 wash buffers

    * Wash head                        1 x 8

    * Dispensing volume                200 - 2500 ㎕/well managed per assay

    * Precision                         ± 3 % CV @ 300 ㎕

    * Residual volume                  <0.5 ㎕ in U-shaped wells,  <2 ㎕ in plat-bottomed wells

    * Buffers level sensor               Yes

    * Waste tank level sensor          Yes

    * Wash cycles                       1 to 9, managed per assay

    * Soak time                         1 to 999 sec. managed per assay

    * Dispensing pressure               Adjustable per assay


    READING UNIT

    * Reading                           vertical with photodiodes, absorbance or kinetics

    * Channels                          8

    * Method                           single, double or double beam with overrange filter

    * Spectrum                         400 - 700 nm

    * Filter                             Up to 8 positions available, 4 already on board

                                         (405, 450, 492, 620 nm)

    * Reading time                     Less than 10 sec

    * Dynamic range                   - 0.100 to 3.500 absorbance units

    * Linearity                         0 - 2.000 absorbance unit ±1.0 %

    * Accuracy                         ±0.005 absorbance unit or 2.5 %


    MANAGEMENT SYSTEM

    * Computer                        Pentium III, 500MHz, 64Mbytes RAM

    * Floppy drive                      3.5"

    * Hard disk                         6.4 GBytes

    * Keyboard                         Alphanumeric

    * Mouse                            Standard

    * Monitor                           15" colour

    * Printer                            Standard Ink jet


    SOFTWARE FEATURES

    * Operating system                 Window 2000 or 95 or 98    32-bit application

    * Language                         Multilanguage

    * Plate capacity                     4 up to 7, in continuous loading

    * Multiple assays per plate         Yes up to 12 assays

    * Data reduction                    Cut-off (qualitative)

                                         Interpolation method (quantitative):4 parameter, point to point,

                                         liner regression, cubic, spline, etc.

    * QA analysis                      mean, SD, CV, Levey-Jennings

    * Protocol storage                  Related to HD capacity

    * Result print out                  Definable per assay and patient

    * Patient archive                   Yes

    * Plate loading                     Per plate, managed with time scheduling

    * Process in control                Yes (on - line log event/error file)

    * I/O interface                     ASTM and Flexible ASCII.


    DIMENSIONS

    * Width                            1130 mm

    * Depth                             760 mm(880mm including the pipette waste bag)

    * Height                           1000 mm

    * Weight                             130 Kg


    ELECTRONIC REQUIREMENTS

    * Universal a.c. input               90 - 260 VAC, 47 - 62 Hz

    * Power                            Typically 500 VA max.


    SAFETY REQUIREMENTS

    * CE mark according to Directives 89/336/EEC and 73/23/EEC